Recall Z-1750-2008 ยท LXH Orthopedic Manual Surgical Instrument

Stryker Howmedica Osteonics Corp.

Omnifit EON Surgical Protocol; Literature Number: LSP48 4/04 Stryker Howmedica Osteonics Corp., Mahwah, NJ

DateAug 17, 2008 (initiated May 17, 2007)
StatusTerminated
ReasonThe Table 2 Broach and Neck Trial Sizing (Literature # LSP48 04/04) on page 3 is incorrect, reading 5mm less than what it should be for certain sizes.
Root causeLabeling design
ActionStyrker Branches/Agencies were notified of the incorrect information printed in the protocal on May 17, 2007 via an Important Product Correction letter. The letter requested that all incorrect copies of the protocol be located and destroyed immediately. They requested that the Acknowledgement Form be completed and faxed, and that a Stryker Orthopaedics Customer Service rep be contacted in order to reorder the destroyed product.
Quantity2501 units
DistributionWorldwide Distribution --- including USA and countries of Canada, Australia, Latin America and Greece.