Recall Z-1623-2008 ยท LZG Pump, Infusion, Insulin

Smiths Medical MD, Inc.

Deltec Cozmo Insulin Pump, Model 1800 Smiths Medical International Ltd WD24 4LG UK, Smiths Medical MD, Inc., St Paul, MN 55112. This Pump is similar to the currently distributed Deltec Cosmo Insulin Pump Model 1700 pump, with the exception of updated software that enhanced certain features and added a food data base to allow users to better manage dietary requirements in conjunction with the use

DateAug 15, 2008 (initiated Mar 10, 2008)
StatusTerminated
ReasonSmith Medical discovered an issue with a motor component in specific serial numbers of the Deltec Cozmo Insulin Pump, Model Numbers 1700 and 1800. This issue affects how the motor operates and may cause an over-delivery of insulin which could result in injury to the user.
Root causeComponent change control
ActionConsignees were mailed an Urgent Medical Device Recall letter dated March 10, 2008. The Urgent letter addressed to "Distributors, Clinicans, Patients and Other Customers" contains Description of the Problem, Results of Investigation, and Advise on Action to be taken by the User. The patient users will be required to return a Recall Confrirmation Form. The USA distributors, USA prescribing physicians and the USA managed Care Accounts (Insurance Companies) will be mailed also a cover letter.
Quantity1066
DistributionClass II Recall - Worldwide Distribution --- USA including states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME. RO, MN, MO, MS, ME, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of Greece, Netherland
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062323DELTEC COZMO INSULIN INFUSION PUMP WITH COZMONITOR BLOOD GLUCOSE METER, MODELS 21-1815, 21-1816, 21-1817
Smiths Medical MD, Inc., Nov 1, 2006