Recall Z-1895-2008 ยท KPR System, X-Ray, Stationary

Philips Medical Systems North America Co. Phillips

Essenta DR - digital multifunctional x-ray system (application SW release 1.0.2 and lower). The device is intended for routine exams, including specialist areas like trauma or pediatric work, excluding mammography.

DateSep 16, 2008 (initiated Apr 8, 2008)
StatusTerminated
ReasonBright artifacts may appear on a patient image from a previous exposure, which might lead to a misdiagnosis.
Root causeSoftware design
ActionOn 4/08/08, the firm notified customers by letter, sent by certified mail. The letter explained the recall and advised customers of the following: "If direct radiation reaches the detector, the collimation should be not too tight, such that the area of direct radiation is larger than 30 mm x 30 mm. (1.2 inch by 1.2 inch). The reason is that a minimum area is needed to detect and prevent memory artifact... All possibly affected products in the field will be upgraded with FCO 71200026 to resolve the issue." Contact Sarah Baxter at 425-487-7665 or Philips Call Center at 800-722-9377, #5, #2 if
Quantity10 units
DistributionNationwide Distribution -- including states of SC, MN, CA, GA, TX, NY, OR and UT.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070528PHILIPS ESSENTA DR
Philips Medical Systems GmbH, Dmc, Mar 9, 2007