Recall Z-1609-2008 ยท FZP Clip, Implantable

Medtronic Cardiac Surgery Technologies

VNUS Medical U-CLIP Removal Tool, UREM1860, is intended for the removal of U-Clip anastomotic devices.

DateSep 16, 2008 (initiated Feb 25, 2008)
StatusTerminated
ReasonMedtronic has identified one lot #0610315 of U-Clip Removal Tools, Catalog UREM1860FW, that was incorrectly assembled. The nose cone attachment, on the body of the removal tool, may not contain sufficient adhesive, which can allow the nose cone to separate from the body of the device.
Root causeProcess control
ActionA Medtronic "Urgent Medical Device Field Notification" letter was sent to one distributor and a separate consignee Medtronic letter "Urgent Medical Device Field Notification" was sent to end users on February 25, 2008. The letters explained the situation and requested that the impacted product be returned to Medtronic for disposal. The end user will be requested to complete a Recall certificate acknowledging the receipt of the information. If you have questions, contact your VNUS Sales Representative or Lauralie Millikan at 763-391-9124.
Quantity80
DistributionAZ, GA, PA, MI, KY, FL, OH, TX, NY, VA, CA, MO IL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031623MODIFICATION TO COALESCENT SURGICAL U-CLIP AND ACCESSORIES
Coalescent Surgical, Jul 3, 2003