Recall Z-1703-2008 · DRF Catheter, Electrode Recording, Or Probe, Electrode Recording

Biosense Webster Inc

Circular Mapping Catheter-Biosense Webster LASSO 2515 Variable Circular Mapping Catheter- The catheter is indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e., recording and stimulation only. The device is designed to obtain electrograms in the atrial regions of the heart.

DateAug 25, 2008 (initiated Mar 25, 2008)
StatusTerminated
ReasonThis recall was initiated after the firm received several complaints reporting issues with deflection. The catheter may become caught in the mitral valve, requiring surgical procedures, valve may be torn while trying to dislodge, unable to retract the catheter through the sheath and require surgical intervention to remove, pulling on catheter to remove may lead to large atrial septal tear, and sep
Root causeDevice Design
ActionA letter entitled, "Urgent Safety Field Notice" dated March 25, 2008, and certification form were sent to all customers and Johnson & Johnson affiliates who purchased the product in the past year as of March 25, 2008, due to the expiration date. For customers in the US, two letters and forms were sent out - one addressed to the EP/Cardiology Lab and a second to the Risk Management Office. If you have questions or have not received a letter contact Customer Service at 1-800-729-9010.
Quantity14,295 units for both products
DistributionNationwide and Worldwide to: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Columbia, Czechia, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary , Iran, Ireland, Isra¿¿l, Italy, Japan, Lithuania, Luxembourg, Mexico, Middle East, Netherlands, Norway, Panama,