Recall Z-1692-2008 ยท CBK Ventilator, Continuous, Facility Use

Covidien Limited

Puritan-Bennett 800 Series Ventilator Backup Power Source (BPS) Model 802, Puritan-Bennett Corporation, A subsidiary of Mallinckrodt Inc., Carlsbad, CA

DateSep 16, 2008 (initiated Mar 21, 2008)
StatusTerminated
ReasonWiring in battery back-up power supply may short and cause thermal damage to ventilator.
Root causeProcess control
ActionAll customers were notified by Federal Express letter and were contacted by phone through field service engineers. They were told to examine their inventory for any installed or spare part battery assemblies and to remove and quarantine these parts. Any ventilators with installed affected battery assemblies were to be taken out of service.
Quantity6 Ventilators and 151 factory replacement units
DistributionNationwide, including one VA facility in IL and two military facilities in NC and NY. No foreign distribution.
FDA recordOpen on fda.gov