Recall Z-1690-2008 ยท LXH Orthopedic Manual Surgical Instrument

Stryker Howmedica Osteonics Corp.

Universal Notch Preparation Guide, Size #3; Non Sterile; Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430; authorized representative in Europe: Stryker France ZAC Satolas Green Pusignan; Cedex, France.

DateSep 20, 2008 (initiated Mar 18, 2008)
StatusTerminated
ReasonDiscovery of the potential for Pin and Punch interference of the Size 3 Scorpio Notch Preparation Guide that may result in damage to bone if a user aggressively attempts to seat an obstructed punch.
Root causeDevice Design
ActionUrgent Product Correction notification letters were sent via Federal Express, return receipt to all users on March 18, 2008. Questions should be directed to Rita Intorella, Divisional Regulatory Reporting at (201) 831-5825.
Quantity739 units in US; 1213 units internationally
DistributionProducts were distributed worldwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.