Recall Z-1642-2008 ยท IXI Block, Beam-Shaping, Radiation Therapy

Philips Medical Systems North America Co. Phillips

AD 7 Patient table supplied with Allura Xper FD10 and Allura Xper FD10/10. Product is non-tilting but can be purchased with a tilt option.

DateSep 1, 2008 (initiated Mar 14, 2008)
StatusTerminated
ReasonPotential for radiologic patient table to become immobile and unable to move again due to force sensor sensitivity to electromagnetic radiation.
Root causeDevice Design
ActionOn 03/22/2004, firm sent letters via certified mail to notify customers of issue. Letter advises customers of potential risk for patients and users, and states to not use system for critical examinations or treatments if table movements suddenly stops or makes a freeing movement. The letter indicates a representative will contact customer for installation of a Hardware filter, free of charge, to prevent Electromagnetic interference (EMI) from affecting the force sensor. Please contact Philips at 1-800-722-9377 #5, #4, #1.
Quantity39 units
DistributionProduct is distributed to hospitals and medical centers nationwide to the following states: ND, FL, KY, MN, VA, ME, PA, NJ, CT, TX, CA, OH, TN, OR, GA, NY, SC, MO, WA, IL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041949ALLURA XPER FD10
Philips Medical Systems North America Co., Jul 30, 2004