Recall Z-2422-2010 ยท KPR System, X-Ray, Stationary

Philips Medical Systems North America Co. Phillips

DigitalDiagnost X-ray System is a stationary Diagnostic X-ray using Digital Solid State Imaging Detector. It is identified with Part Number 4512 201 04752 Software Version 1.5.1 and Part Number 4512 201 04753 Software Version 1.5.2. The device is manufactured by Philips Medical Systems DMC GmbH, RontgenstraBe 24, 22335 Hamburg, Germany.

DateSep 14, 2010 (initiated Dec 13, 2007)
StatusTerminated
ReasonWhen using DICOM print functionality, the printed images may contain data for the wrong patient.
Root causeSoftware design
ActionPhilips Healthcare issued an "URGENT PRODUCT CORRECTION " letter dated March 10, 2008 to their consignees. The letter described the product, problem and action to be taken by the customer. The letter instructed the customer to: 1) Validate that image and patient information are correct on hard copy prints. 2) Reprint images that have mixed patient information, and re-check that image and patient information are correct. 3) Do not use printed images if the information is incorrect. The firm will issue the Field Change Order (FCO) 71200025 and the field service engineers will visit each consig
Quantity119 units
DistributionNationwide Distribution: including the states of AR, AZ, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, PA, SC, TN, TX, UT, VT, WA, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K982795PHILIPS BUCKY VISION
Philips Medical Systems, Inc., Nov 24, 1998