Recall Z-1603-2008 ยท FNL Bed, Ac-Powered Adjustable Hospital
Hill-Rom, Inc.
Hill-Rom 100 Low Bed; Model No. P3930.
| Date | Sep 5, 2008 (initiated Apr 8, 2008) |
| Status | Terminated |
| Reason | The upper deck may collapse to its lowest position and the caster brakes may fracture if the bed is moved while the brakes are locked. |
| Root cause | Device Design |
| Action | Consignees were notified via letter dated April 8, 2008 to lower these beds to their lowest position, to remove the beds from use if they move more than 1/2 inch with the brakes locked and that the firm will repair these beds when fixes become available. If you have questions regarding this recall, contact Hill-Rom Technical Support at 800-445-3730. |
| Quantity | 753 beds |
| Distribution | Worldwide: USA and Canada. |
| FDA record | Open on fda.gov |