Recall Z-1558-2008 ยท GCJ Laparoscope, General & Plastic Surgery

Stryker Endoscopy

24mm Rotating C-Mount Coupler, Model Number 1088-020-121S2. The device in indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery wherever a laparoscope, endoscope or arthroscope is indicated for use.

DateAug 24, 2008 (initiated Mar 31, 2008)
StatusTerminated
ReasonThe coupler on this device was manufactured without set-screws that are used to hold and center the rear assembly and the endobody. The coupler is part of the camera system and affects the ability to focus.
Root causeNonconforming Material/Component
ActionOnly one customer had all of the defective couplers. The firm called this customer on March 27, 2008, and asked that the 28 couplers remaining in their possession be returned.
Quantity28 units
DistributionProduct was delivered to one consignee in Iowa.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K9835663-CHIP VIDEO CAMERA
Stryker Endoscopy, Jan 11, 1999