Recall Z-1764-2008 ยท LGC Enzyme Linked Immunoabsorbent Assay, Herpes Simplex Virus, Non-Specific

Focus Diagnostics Inc

HerpeSelect 1 and 2 Immunoblot IgG Kit, IB0900G Intended for qualitatively detecting the presence or absence of human IgG class antibodies to HSV-1 and HSV-2 in human sera.

DateAug 17, 2008 (initiated Jul 6, 2007)
StatusTerminated
ReasonA small percentage of samples tested in specific lots, the HSV common antigen band appears lighter (less reactive) than the reading control band. Reduced reactivity of the common antigen band may lead to inaccurate interpretation of a sample as negative.
Root causeProcess design
ActionFocus notified all clients which received the recalled lots starting on July 6, 2007 via an Important Medical Device Information letter by fax or email. The letter explains how to identify suspect results and the steps to be taken if suspect results are obtained. If customers find the interpretation of results with the indicated kit lots to be problematic, customers have the option of having their kits replaced at no charge within 90 days of receiving the notification.
QuantityLot Code 063169 - 97 Kits, Lot Code 063354 - 16 Kits, Lot Code 063355 - 74 Kits
DistributionWorldwide Distribution --- USA including states of OH, CA, CT, IL, NY, and countries of Barbados, Canada, Germany, Israel, Slovakia, France, and England.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K000238HSV-1 & HSV-2 DIFFERENTIATION IMMUNOBLOT IGG, MODEL IB0900G
Mrl Diagnostics, Apr 14, 2000