Recall Z-0869-2008 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Roche Diagnostics Corp.

Roche/Hitachi Modular Analytics System, Modular D Module GMMI clinical chemistry analyzer; Catalog number 03284522001. Intended for qualitative and quantitative in vitro determinations using a wide variety of tests. Roche Diagnostics Corporation, 9115 Hague Road, Indianapolis, IN, 46250

DateAug 13, 2008 (initiated Mar 21, 2008)
StatusTerminated
ReasonPossible mismatch between patient and result.
Root causeProcess change control
ActionConsignees were notified via Urgent Medical Device Correction letter dated 3/21/08, which instructed them to not manually order QC test selections while the system is in operation mode and that they can only manually order QC when the system is in Stand-by mode. Consignees are instructed to consult with their physicians or pathologist to determine specific clinical implications for their patients. The firm states that this issue will be resolved in the next version of software.
Quantity11 of all models
DistributionNationwide Distribution --- including states of Florida, Georgia, Indiana, New Jersey and North Carolina.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K953239BOEHRINGER MANNHEIM/HITACHI 917 ANALYZER
Boehringer Mannheim Corp., Sep 25, 1995