Recall Z-1590-2008 ยท LZG Pump, Infusion, Insulin
Smiths Medical MD, Inc.
Deltec Cozmo Insulin Pump, model 21-1700. A syringe infusion pump designed for Continuous Subcutaneous Insulin Infusion for the control of diabetes. Reorder Numbers: 21-1701-02/32/33/36/44/49/51/54/62/63/66/81 21-1702-02//49/51 21-1703-02/32/33/36/44/49/51/54/62/63/66/81 21-1704-02/32/33/36/44/49/51/54/62/63/66/81 21-1705-79 21-1706-79 21-1707-79 21-1711-01/02/03/06/14/50 21-1712-01/02 21-1713-01
| Date | Sep 1, 2008 (initiated Feb 22, 2008) |
| Status | Terminated |
| Reason | Smith Medical has received reports of Deltec Cozmo Insulin Pumps powering down without an alarm. Investigation concluded that if the battery cap is not fully tightened, the Pump may power down and a brief "chirp" will sound. This may occur if the user has not sufficiently tightened the battery cap or if the battery cap is damaged. If a patient is asleep or does not regularly check their blood g |
| Root cause | Device Design |
| Action | Consignees and the USA demonstration pump physicians were mailed an "Urgent Medical Device Correction" letter on 2/22/08. The letter stated Details on Affected devices, Description of the problem, Advice on action to be taken by the user, Additional Information and New Warning. These consignees will be required to return the Correction Confirmation Form. The 17 countries with international mailing addresses were sent the Urgent Medical Device Correction letter (Urgent Field Safety Notice) via the Smiths Medical International Regulatory department based in Watford UK. The Smiths Medical Dist |
| Distribution | Worldwide Distribution --- USA including states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of AU, CA, CL, CN, DE, EG, |
| FDA record | Open on fda.gov |