Recall Z-1589-2008 ยท JPD Hemoglobin A2 Quantitation

Bio-Rad Laboratories Inc

VARIANT -thalassemia Short Program Hemoglobin Testing System, Model Number 270-0002 (500 tests), with -thalassemia ROM card lot PA 62600, PIN 2700104

DateSep 11, 2008 (initiated Mar 18, 2008)
StatusTerminated
ReasonSome product ROM cards have an incorrect program loaded and are unable to be updated.
Root causeMixed-up of materials/components
ActionThe firm initiated its recall on March 18, 2008 by issuing a recall notice to customers and subsidiaries via e-mail. Not all VARIANT -Thalassemia Short ROM Cards are affected. Inventory should be examined and quarantine the product subject to this Recall. Contact your regional Bio-Rad office for ROM Card replacement or if you have questions.
Quantity64 kits total
DistributionProduct was distributed to 17 consignees and 7 subsidiaries Worldwide. Distribution included to states of CA, TX, WI, VA, UT, PA, TN, OH, FL, MA, OR, AR, IL, MD, CO, and NY, and countries of UK, Australia, Canada, France, Singapore, New Zealand, Italy.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K924122VARIANT BETA-THALASSEMIA SHORT PROGRAM