ClearanceBrief
Recall Z-1488-2008 ยท GCY Apparatus, Suction, Single Patient Use, Portable, Nonpowered

Zimmer Inc.

Zimmer Hemovac Wound Drainage Device, 400ml Comp Bonded Evac 3/16" PVC w/trocar, 4,8mm, 15Fr, PVC, 107cm, soft, sterile, latex free, Rx only; Zimmer, Swindon, Wiltshire, United Kingdom; Zimmer, Dover, Ohio; REF 00-2505-700-10.

DateSep 19, 2008 (initiated Jan 29, 2008)
StatusTerminated
ReasonThe sterility of the device may be compromised due to the possible presence of slits or pinholes in the packaging.
Root causePackaging process control
ActionConsignees were notified via recall letter dated 1/29/08 to locate and return the affected product. The recall action was expanded to additional products/lots via recall letter dated 3/17/08. Custom kit packers were instructed to notify their customers.
DistributionNationwide, Australia, Canada, Chile, China, Czech Republic, Dubai, Ecuador, France, Germany, Greece, Hungary, Italy, Japan, Korea, Poland, Romania, Russia, Saudi Arabia, Serbia & Montenegro, Singapore, Slovakia, South Africa, Spain, and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.