Recall Z-0561-2010 ยท JAK System, X-Ray, Tomography, Computed

Xoran Technologies, Inc.

Xoran MiniCAT System for ENT (server and work station), Xoran Technologies, Ann Arbor, MI. Cone beam CT for sinus, temporal bones, and maxillofacial imaging.

DateDec 23, 2009 (initiated Sep 17, 2008)
StatusTerminated
ReasonThe system may overheat, causing poor image quality, interruption in image acquisition and/or damage to the system. Also, if the system is not allowed to cycle down prior to being restarted after an emergency stop, the x-ray and gantry motions will both resume.
Root causeDevice Design
ActionA letter dated 9/17/08 was sent to each account telling them that the PLC would be upgraded. The firm visited each account and made corrections.
Quantity82
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032243MINICAT DENTOCAT