Recall Z-1633-2008 ยท DQY Catheter, Percutaneous

Abbott Vascular-Cardiac Therapies dba Guidant Corporation

6F Viking Guiding Catheter, Percutaneous catheter, Shape ALR 1.2, Part Number: 1001934-06, Abbott Vascular, Santa Clara, CA 95054. Designed to provide a pathway trough which therapeutic and diagnostic devices are introduced.

DateAug 13, 2008 (initiated Mar 17, 2008)
StatusTerminated
ReasonMislabeling: The part number on the product label packaging may not match the product.
Root causeLabeling mix-ups
ActionConsignees were notified via an Urgent Device Recall letter dated and hand delivered starting 3/18/08. The letter requests users to remove and quarantine the affected lots, which will be picked up and replaced by the recalling firm. Users were also requested to return the enclosed Efficiency Check Reconciliation Form. Abbott Vascular will exchange all returned devices. For additional information contact 1-800-227-9902.
Quantity10 units
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K972484ACS VIKING GUIDING CATHETER
Guidant Corp., Sep 26, 1997