ClearanceBrief
Recall Z-1711-2008 ยท FRN Pump, Infusion

Cardinal Health 303 Inc DBA Alaris Products

Grey Inter-Unit Interface Connectors ("IUI") on the following Alaris System modules built on or before March 26, 2004: (1) Alaris PC point-of-care unit ("PC Unit") (Model 8000)* (a.k.a. Medley PC Unit), (2) Alaris Pump Module (Model 8100) (a.k.a. Medley Pump Module), (3) Alaris Syringe Module (Model 8110) (a.k.a. Medley Syringe Module), (4) Alaris PCA Module (Model 8120) (a.k.a. Medley PCA Mod

DateJun 10, 2008 (initiated Mar 5, 2008)
StatusTerminated
ReasonA channel error on an affected Alaris System module will stop an active infusion (or monitoring) with an audible and visual alarm. Channel errors with the Alaris System due to the Inter-Unit Interface (IUI) connectors on all Alaris System modules (manufactured on or before March 26, 2004). Grey Inter-Unit Interface Connectors ("IUI") on the following Alaris System modules built on or before March
Root causeDevice Design
ActionAn Urgent Medical Device Recall Notification letter was sent to consignees on April 10, 2008. Four customer letters were issued. The U19 IC issue affects only the Pump module and the U9 IC issue affects only the Syringe and PCA modules. Therefore there will be one letter for customers affected by all three issues Ul9 IC, U9 IC, and the IUI, one letter for affected Ul9 IC and U9 IC customers, one letter for U19 IC-only customers and one letter for U9 IC-only customers. All mailings were conducted through consignees (distributors as well as direct customers for the affected devices). Distrib
Quantity77,515 Units
DistributionClass II Recall - Worldwide Distribution --- USA including states of AR, AZ, CA, CT, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, ME, MD MI, MN, MO, MS, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI and WY, Guam and country of Canada. ****CDRH Comments: "Note for District Off
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010966MEDLEY SP02 MODULE, MODEL 8220
Alaris Medical Systems, Inc., Jun 8, 2001
K012383GEMINI PC-1,PC-2, PC-2TX,PC-4, SIGNATURE EDITION INFUSION PUMP (SE), MEDSYSTEM III INFUSION (MSIII), MEDSYSTEM PATIENT C
Alaris Medical Systems, Inc., Oct 2, 2001
K022677MEDLEY SPO2 MODULE, MODEL 8210 & SPO2 SENSORS & CABLE
Alaris Medical Systems, Inc., Sep 11, 2002
K023264MEDLEY SYRINGE PUMP MODULE SYSTEM, MODEL 8110
Alaris Medical Systems, Inc., Dec 19, 2002
K043299MEDLEY PCA MODULE, MODEL 8120
Alaris Medical Systems, Inc., Dec 10, 2004
K950419THE IMED ORION INFUSION PUMP AND ADMINISTRATION SETS
Imed Corp., Jun 21, 1995