Recall Z-1710-2008 ยท FRN Pump, Infusion

Cardinal Health 303 Inc DBA Alaris Products

U19 Socketed Integrated Circuit ("U19 IC") on the logic board of Alaris Pump modules (aka. Medley Pump modules), Model 8100 built on or before September 8, 2005. CardinalHealth, San Diego, CA 92130

DateJun 10, 2008 (initiated Mar 5, 2008)
StatusTerminated
ReasonA channel error on an affected Alaris System module will stop an active infusion (or monitoring) with an audible and visual alarm. Channel errors due to failures of the U19 Socketed Integrated Circuits (U19 IC) on the logic board of the Alaris Pump module ("Pump module"). The user will be required to replace the affected module.
Root causeDevice Design
ActionAn Urgent Medical Device Recall Notification letter was sent to consignees on April 10, 2008. Four customer letters were issued. The U19 IC issue affects only the Pump module and the U9 IC issue affects only the Syringe and PCA modules. Therefore there will be one letter for customers affected by all three issues Ul9 IC, U9 IC, and the IUI, one letter for affected Ul9 IC and U9 IC customers, one letter for U19 IC-only customers and one letter for U9 IC-only customers. All mailings were conducted through consignees (distributors as well as direct customers for the affected devices). Distrib
Quantity85,286 units
DistributionClass II Recall - Worldwide Distribution --- USA including states of AR, AZ, CA, CT, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, ME, MD MI, MN, MO, MS, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI and WY, Guam and country of Canada. ****CDRH Comments: "Note for District Off
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012383GEMINI PC-1,PC-2, PC-2TX,PC-4, SIGNATURE EDITION INFUSION PUMP (SE), MEDSYSTEM III INFUSION (MSIII), MEDSYSTEM PATIENT C
Alaris Medical Systems, Inc., Oct 2, 2001
K950419THE IMED ORION INFUSION PUMP AND ADMINISTRATION SETS
Imed Corp., Jun 21, 1995