Recall Z-1453-2008 ยท GCY Apparatus, Suction, Single Patient Use, Portable, Nonpowered

Zimmer Inc.

Zimmer Hemovac Wound Drainage Device Infection Control Kits, Catalog Number/ REF 00-2550-702-10; 400 ml INF. CNTRL. Kit W/ 1/8" Drain Spec., 3.2mm, 10Fr., PVC, 107cm, medium, sterile, latex free, Rx only; Sold in packages of 10, which contain packages of 1 and 5; Zimmer U.K. Ltd., Swindon, Wiltshire, United Kingdom; Zimmer, Dover, Ohio.

DateApr 9, 2008 (initiated Jan 31, 2008)
StatusTerminated
ReasonLeakage: The kits may disassemble at the fluid collection port, which would present a risk of exposure to blood borne pathogens to health care providers should it occur during use.
Root causeProcess control
ActionConsignees were notified via Urgent Infection Control Recall letter dated 1/31/08 to examine their inventory and return the recalled lots. Custom kit packers were instructed in the same letter to recall these products/lots from their own customers.
Quantity34,354 (Recalls Z-1448/1458-2008)
DistributionWorldwide-USA, Australia, Canada, Colombia, Denmark, Dubai, Italy, Japan, Mexico and Singapore.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.