Recall Z-1658-2008 ยท KWT Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented

Biomet, Inc.

Biomet Hybrid Glenoid Boss Cutter; item 406150. - Surgical instrument used to prepare the natural shoulder glenoid surface for implanting a Hybrid Glenoid component. This instrument is intended to function by drilling a hole into the natural glenoid surface allowing the center boss of glenoid implant to sit within the hole, and the non-articulating surface of the glenoid implant to rest flush ont

DateSep 16, 2008 (initiated Mar 13, 2008)
StatusTerminated
ReasonThe glenoid implant may not seat fully on the face of the natural glenoid due to inadequate measurement tolerances (tolerance stack-up) in the design of the cutter. Use of the instrument may lead to incomplete seating of the glenoid implant, elevating the theoretical risk of glenoid loosening, which may lead to early revision surgery. The problem may contribute to early device failure and patien
Root causeDevice Design
ActionConsignees were notified by Urgent Medical Device Removal Notice dated 3/13/08 to discontinue use of the instruments and return to the firm. Operating room managers were sent a separate letter dated 3/13/08 and informed of the possible effects to patients from previous use of this instrument and it was suggested they monitor patients for these effects. If you have any questions contact Mary Hardesty at 1-800-348-9500 or (574) 372-3983.
Quantity189
DistributionClass II Recall - Nationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060694MODULAR HYBRID GLENOID
Biomet Manufacturing Corp, Dec 11, 2006