Recall Z-0274-2008 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

GE Healthcare

Precision 500D Classical R&F System with control room PC; Image-intensified fluoroscopic x-ray system; ; Model Number: 5117866-27.

DateAug 2, 2008 (initiated Aug 1, 2007)
StatusTerminated
ReasonInaccurate cassette size display reading/ Inaccurate mAs reading: 1. When the device is in PBL mode and a larger film cassette is inserted into the table Bucky (replacing a smaller cassette) after a protocol is selected on the console, the collimator will not adjust to the larger cassette size. The collimator stays at the original smaller cassette size and allows the operator to take an exposure.
Root causeSoftware design
ActionConsignees were notified by two "Product Safety Notification" letters on July 10, 2007. Each letter addressed one of the two safety issues with the affected product. The letters provided safety instructions on how avoid the software error and informed users that their representative would contact them to schedule a visit for a software correction installation.
Quantity101
DistributionWorldwide Distribution --USA including states of: AK, AL, AR, CA, CO, CT, FL, GA, IA, ID, IL ,IN, LA, MA, MD, MI, MO,MS, NC, ND, NE, NH, NJ, NM, NV, OH,OK, OR, PA, SC, SD, TN, TX, UT, VA, VT and WA; USA territory of: PR and countries: EGYPT, SAUDI ARABIA, MACEDONIA, FRANCE and GREAT BRITAIN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011624EXPEDIO 500D R&F X-RAY SYSTEM
GE Medical Systems, Jun 8, 2001