Recall Z-1758-2008 ยท IYE Accelerator, Linear, Medical
Siemens Medical Solutions USA, Inc
Siemens Medical Solutions, Mevatron M2/ Primus Mid Digital Linear Accelerator, part number 1940035 equipped with a Digital Electron Variable Applicator (DEVA, Part Number 8485971) with available energy of 6 MeV.
| Date | Sep 29, 2008 (initiated Dec 10, 2007) |
| Status | Terminated |
| Reason | Radiation leak: When used with field sizes of 5cm x 5cm or smaller, product may leak radiation at a distance of 2 cm from the side of the applicator body up to 13%, in excess of IEC standards. |
| Root cause | Device Design |
| Action | Siemens sent a Customer Safety Advisory Notice dated 12/12/2007 to DEVA customers. The letter stated that Siemens would temporarily disable energies 5MeV and below for all customers that decide to continue using the DEVA, resulting in the temporary unavailability of energies 5MeV and below for all available. Customers wishing not to have these energies blocked must return their DEVA and they would be contacted by a local Siemens rep to schedule a service appointment. The letter stated that as a preventative measure that those customers should refrain from using the DEVA with any combination |
| Quantity | 168 total |
| Distribution | Worldwide Distribution: USA, Canada, Belgium, Bulgaria, France, Germany, Hungary, India, Italy, Japan, Malaysia, P.R. China, Pakistan, Poland, Portugal, Qatar, Republic Korea, South Africa, Spain, Sweden, and United Kingdom. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K862339 | MEVATRON KD-2 Siemens Medical Laboratories, Inc., Oct 29, 1986 |