ClearanceBrief
Recall Z-1755-2008 ยท IYE Accelerator, Linear, Medical

Siemens Medical Solutions USA, Inc

Siemens Medical Solutions, Mevatron KD-2 Digital Linear Accelerator, part number 9822685 equipped with a Digital Electron Variable Applicator (DEVA, Part Number 8485971) with available energy of 6 MeV. Medical charged-particle radiation therapy system.

DateSep 29, 2008 (initiated Dec 10, 2007)
StatusTerminated
ReasonRadiation leak: When used with field sizes of 5cm x 5cm or smaller, product may leak radiation at a distance of 2 cm from the side of the applicator body up to 13%, in excess of IEC standards.
Root causeDevice Design
ActionSiemens sent a Customer Safety Advisory Notice dated 12/12/2007 to DEVA customers. The letter stated that Siemens would temporarily disable energies 5MeV and below for all customers that decide to continue using the DEVA, resulting in the temporary unavailability of energies 5MeV and below for all available. Customers wishing not to have these energies blocked must return their DEVA and they would be contacted by a local Siemens rep to schedule a service appointment. The letter stated that as a preventative measure that those customers should refrain from using the DEVA with any combination
Quantity168 total
DistributionWorldwide Distribution: USA, Canada, Belgium, Bulgaria, France, Germany, Hungary, India, Italy, Japan, Malaysia, P.R. China, Pakistan, Poland, Portugal, Qatar, Republic Korea, South Africa, Spain, Sweden, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K862339MEVATRON KD-2
Siemens Medical Laboratories, Inc., Oct 29, 1986