ClearanceBrief
Recall Z-1397-2008 ยท IYE Accelerator, Linear, Medical

Siemens Medical Solutions USA, Inc

Oncor Impression Plus Medical Linear Accelerator, Material Number : 5857912, with control console software versions 6.5/7.2, 8.0, 9.0 and 9.1, Manufactured by Siemens Medical Solutions USA, Inc, Concord, CA

DateApr 2, 2008 (initiated Jan 11, 2008)
StatusTerminated
ReasonLinear accelerator gantry may rotate in an unexpected direction and may cause patient injury on impact.
Root causeSoftware design
ActionThis is a mandatory software upgrade for consignees. The firm's representatives have contacted consignees via Urgent Device Notification letters dated February 14, 2008, notifying them of the new update and making arrangements to install it.
Quantity187 units
DistributionWorldwide-USA including states of AL, AK, AZ, AR, CA, CO, DE, DC, FL, GA, IL, IN, KS, KY, LA, MD, ME, MI, MN, MS, MO, MN, NE, NV, NH, NJ, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, VT, VA, WA, WV, WI, and WY and countries of Angola, Australia, Austria,Belgium,Brazil,Canada,Chile,Colombia,Costa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K993425PRIMUS
Siemens Medical Solutions USA, Inc., Nov 26, 1999