Recall Z-1444-2008 ยท LLZ System, Image Processing, Radiological

Philips Healthcare Informatics, Inc.

Philips Medical Systems iSite PACS (Picture Archiving and Communications Systems) version 4.1.17.0 - Radiology Image management system.

DateAug 16, 2008 (initiated Jan 29, 2008)
StatusTerminated
ReasonWhen using the Freehand Region of Interest (ROI) tool, errors may occur in area calculation yielding incorrect results. The firm discovered that a descrepancy exists in the measurement provided by the tool.
Root causeSoftware design
ActionThe firm issued product notification letter February 20, 2008, to its consignees. The notification alerts customers to the behavior, requires them to post the notification and notify users not to use the Freehand ROI feature until they are upgraded to a corrected software version. Philips will deliver a software upgrade to correct the problem.
Quantity2 units
DistributionWorldwide Distribution --- USA including states of MN, CO, and PA, and country of Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063267ISITE PACS, V 4. X
Stentor Inc., A Philips Medical Systems Co., Nov 9, 2006