Recall Z-1657-2008 ยท HRX Arthroscope

ArthroCare Corporation

ArthroCare brand Topaz XL ArthroWand with Integrated Cable Wand, Catalog Number: AC4045-01; Product is manufactured and distributed by ArthroCare Corporation, Sunnyvale, CA - Sterile Surgical Device. Product is indicated for ablation and debridement of tendons in orthopedic procedures.

DateSep 16, 2008 (initiated Jan 31, 2008)
StatusTerminated
ReasonDevices for which sterility may be compromised as evidenced by punctures in the plastic tray during shipping. The risk to the patient is a potential infection subsequent to the surgical procedure.
Root causePackaging process control
ActionUrgent Product Recall letters (Dated January 31, 2008) were issued to end users of the device (medical professionals) and direct distributors. The letter instructs users to return any affected products.
Quantity2,129
DistributionWorldwide Distribution --- USA including states of AL, AR, AZ, CA, CO FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, ME, MN, MO, NC, ND, NJ, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, WI, and countries of Sweden, Australia, Germany, Italy, UK, Austria, Saudi Arabia, Belgium, Portugal, UAE, Kuwait, Singapo
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K053567ATHROCARE TOPAZ ARTHROWANDS