Recall Z-1541-2008 ยท DXJ Display, Cathode-Ray Tube, Medical

Philips Medical Systems

IntelliVue Clinical Information Portfolio Critical Care Release D, Version D.00.00 Product Number: 865047. This device is a clinical information system intended for use in data collection, storage, and management with independent bedside devices, and other systems that are connected either directly or through networks. It is used when there is a need for generating patient records and computatio

DateAug 23, 2008 (initiated Mar 4, 2008)
StatusTerminated
ReasonMedications prescribed for one patient were printed on the record of another patient.
Root causeSoftware design
ActionPhilips Medical issued letter notification, dated 29 Feb 2008 to customers titled; "Urgent Medical Device Correction Notice". Customers were asked to not solely rely on the Worklist to remind them when to administer orders or interventions until the software has been upgraded. Customers should review the reminders and verify orders and interventions on the Medication Administration Record and Administration record. The correction will be in the form of a software upgrade and will be performed remotely by Philips service personnel.
Quantity138 for products
DistributionNationwide Canada, Australia, Austria, Belgium, Denmark., France, Finland, Germany, Great Britain, Italy, Malaysia, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, and UAE
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K992636HP CAREVUE 9000 CLINICAL INFORMATION SYSTEM, MODEL M2331A
Hewlett-Packard Co., Aug 31, 1999