Recall Z-2003-2008 ยท CBK Ventilator, Continuous, Facility Use

Respironics California Inc

Esprit Power Supply Snubber Board Assembly FRU, PN 1036236 The Esprit Ventilator is a microprocessor controlled, electrically powered, mechanical ventilator.

DateSep 17, 2008 (initiated Jan 25, 2008)
StatusTerminated
ReasonThis action is being taken to address power supply failures on the Esprit Ventilator which have occurred in some units. These failures, which were discovered through routine product monitoring, have been reported in countries that utilize an input voltage of greater than 200 volts AC. Some customers have reported a burning smell coming from the ventilator.
Root causeDevice Design
ActionOn January 25, 2008, letters were sent to all existing Esprit distributors in > 200 volt countries. This letter notified customers of the pending field action, the reason for the action and included a set of instructions that end users need to follow in order to be able to continue using their Esprit pending the installation of replacement parts. The letter also stated that as soon as replacement parts are available, customers will be provided with a serial number list of their affected units. Respironics will provide replacement parts at no cost and instructions to upgrade the Esprit Ventilat
Quantity248 units
DistributionInternationally to: Argentina, Turkey, Peru, Russia, Uruguay, Malaysia, Lebanon, India, Spain, Hong Kong, Brazil, UAE, Cambodia, Thailand, Chile, Philippines, South Korea, Vietnam, Bangladesh, Singapore, Italy, United kingdom, South Africa, Egypt, Pakistan, Jordan, Algeria, and Japan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K001208ESPIRT
Respironics, Inc., May 12, 2000
K023350ESPRIT VENTILATOR WITH RESPIRATORY MECHANICS
Respironics California, Inc., Nov 6, 2002
K034040ESPRIT VENTILATOR, MODEL V1000
Respironics, Inc., Jan 21, 2004
K041412ESPRIT VENTILATOR WITH NICO-ESPRIT INTERFACE OPTION, MODEL V1000
Respironics California, Inc., Jun 18, 2004
K981072ESPRIT VENTILATOR
Respironics, Inc., Nov 12, 1998