ClearanceBrief
Recall Z-1380-2008 ยท BYE Attachment, Breathing, Positive End Expiratory Pressure

Smiths Medical ASD, Inc.

Smiths Medical Oxy-PEEP High Flow O2 with PEEP; Positive end expiratory pressure breathing attachment; Reference Number: P63000; Smiths Medical, Rockland, MA 02370.

DateJul 16, 2008 (initiated Mar 4, 2008)
StatusTerminated
ReasonIncorrect Oxygen Level: Device may not provide expected oxygen concentration or the expected flow rate due to incorrectly molded part. Patient will not receive proper amounts of air/oxygen when in need.
Root causeNonconforming Material/Component
ActionConsignees were notified on 2/26/08 via telephone to place the affected products on hold pending further instruction from the recalling firm. A follow-up letter was issued on 3/4/08 advising consignees to return any product in stock for a replacement or credit. Distributors were requested to provide their customer lists to the recalling firm in order to notify their customers of this recall.
Quantity196
DistributionWorldwide Distribution including USA states of CA, FL, IN, KS, MN, MO, NC, NH, NY, PA and TN, and country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K854822LDS PEEP VALVE
Life Design Systems, Inc., Feb 24, 1986