Recall Z-1460-2008 ยท KPQ System, Simulation, Radiation Therapy

Varian Medical Systems Oncology Systems

Acuity with Conebeam Computed Tomography; Radiation therapy simulation system; Model Number: H77.

DateJul 30, 2008 (initiated Jan 3, 2008)
StatusTerminated
ReasonIncorrect Image Orientation: The image orientation tag may not be set correctly if the patient has been scanned with an orientation other than Head First Spine. When such images are imported to a treatment planning system, they may appear with either a mirrored or a rotated view.
Root causeSoftware design
ActionConsignees were notified by an Urgent Medical Device Correction letter issued on 02/05/2008. The letter provided recommended actions for users to follow in order to avoid the software error. The letter informed users that a service pack is being created to correct the issue and that the recalling firm will notify users when it becomes available. For additional information, contact 888-827-4265.
Quantity16 UNITS
DistributionWorldwide Distribution--USA including states of CA, MA and PA, and countries of UK, Chile, Belgium, Canada and Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033339ACUITY WITH CONEBEAM COMPUTED TOMOGRAPHY
Varian Medical Systems, Inc., Dec 30, 2003