Recall Z-1567-2008 ยท MKJ Automated External Defibrillators (Non-Wearable)

Physio Control, Inc.

FASTPAK batteries labeled Manufactured for Medtronic Physio-Control 11811 Willows Rd, Redmond, WA USA. Batteries are Nickel-Cadmium (NiCd) battery packs used to provide DC power to LIFEPAK 5, 10, 11, or 12 defibrillator/monitors. The FASTPAK battery has a capacity of 1 amp/hour (1.0AH). The FASTPAK battery can be charged in a Battery Support System, Battery Support System 2, Mobile Battery Se

DateAug 24, 2008 (initiated Feb 27, 2008)
StatusTerminated
ReasonBatteries mislabeled - batteries labeled as 2.4 amp/hour batteries, however are 1.0 amp/hour batteries.
Root causeLabeling mix-ups
ActionOn February 27, 2008, the firm sent a "URGENT MEDICAL DEVICE - ACCESSORY RECALL" letter via certified return receipt to their consignees. The letter states the batteries are fully functional 1.0 amp/hour batteries mislabeled as 2.4 amp/hour batteries and that the batteries will provide half the monitoring time and half the number of defibrillation shock as of a 2.4 amp/hour LIFEPAK NiCd battery. Enclosed with the letter is a confirmation sheet for the consignee to indicate the number of recalled batteries they have and fax back to the firm. The firm will send out replacement batteries and th
Quantity811 batteries
DistributionBatteries were distributed nationwide and to Canada, Netherlands, Asia Pacific and Latin America.