Recall Z-1656-2008 ยท MKJ Automated External Defibrillators (Non-Wearable)

Physio Control, Inc.

LIFEPAK 1000 defibrillator - portable battery powered semi-automated model that can be operated in either of three modes: Automated external defibrillator (AED), manual defibrillator and electrocardiographic (ECG) mode. The defibrillator is intended for indoor or outdoor use in a wide variety of hospital and pre-hospital settings including ground and air ambulances, physician's offices, aircraft,

DateSep 16, 2008 (initiated Feb 27, 2008)
StatusTerminated
ReasonPotential for the display screen to dim and eventually go blank. The risk to patient is a delay in defibrillation therapy if the defibrillator was needed in a cardiac arrest situation. Death or permanent impairment can occur if the user fails to deliver a defibrillation shock or defibrillation is delayed.
Root causeComponent change control
ActionOn February 27, 2008 the firm sent a URGENT MEDICAL DEVICE NOTIFICATION letter to consignees advising of the issue. The letter states the firm is performing accelerated function testing and that a follow up notification will be sent if the firm's final test results indicate further action is necessary. The letter recommends consignees "Continue to use your LIFEPAK 1000 defibrillator/monitor Set your device to Auto-Analyze Contact Physio-Control immediately if your display begins to dim or goes blank. What to do if the display goes blank while in use: 1. Continue to use the device to tre
Quantity1,647 total - 426 US, 1,221 international
DistributionWorldwide Distribution --- including USA and countries of Netherlands, Germany, Australia, Hong Kong, Barbados and Brazil.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042404LIFEPAK, MODEL 1000
Medtronic Emergency Response Systems, Inc., Dec 22, 2005