Recall Z-1316-2008 ยท LMB Device, Digital Image Storage, Radiological

Stryker Endoscopy

Stryker Digital Capture System Ultra (SDC Ultra), Model Number 240-050-988, Stryker Endoscopy, image and video archiving system.

DateSep 11, 2008 (initiated Feb 15, 2008)
StatusTerminated
ReasonPatient Identification numbers and names may be applied to images from another patient due to a software issue. This failure mode can only occur when the pre-load function of the SDC ultra is used.
Root causeSoftware design
ActionStryker Endoscopy mailed Urgent Device Correction notices, dated February 14, 2008, to all consignees, and sales representatives will be supplied with a fix to distribute to the user base.
Quantity814 units distributed
DistributionWorldwide distribution: USA, Canada, Greece, and Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.