Recall Z-1472-2008 ยท CGA Glucose Oxidase, Glucose

Nova Biomedical Corporation

Varta Easypack XL Lithium Polymer Batteries (P/N 41772 and 42215) Used with: Nova Biomedical StatStrip Glucose Hospital Meter (P/N 42224) and Nova Biomedical StatSensor Creatinine Hospital Meter and EZ CHEM Creatinine Meter (P/N 9600). Statstrip Meter an In vitro diagnostic for use by health care professionals in the quantitative determination of glucose (GLU) in whole blood. EZ Chem In vitro

DateSep 23, 2008 (initiated Feb 21, 2008)
StatusTerminated
ReasonLithium batteries may fail if the glucose or creatinine meters have been dropped.
Root causeComponent design/selection
ActionNova Biomedical notified consignees by telephone and followed by the Recall Notification letter for the Vata Easy Pack XL Lithium Polymer Battery, dated 2/2108 and mailed 2/26/08. Users are notified of the problem with the battery in the Nova Stat Strip Blood Glucose Hospital Meter System and is replacing batteries by on site visit. E-Z-EM , Inc. notified their customers by letter dated 3/17/08. Contact Nova Biomedical Corporation at 1-800-545-6682 for assistance.
Quantity4432 batteries
DistributionWorldwide Distribution --- including USA and countries of Canada, Taiwan, Hungary, Malaysia, Australia, Russia, Thailand, New Zealand, Austria, France, Germany, UK, China, South Africa, Croatian, Peru, and Czech Republic.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060345NOVA STATSTRIP GLUCOSE HOSPITAL METER, STATSTRIP CONTROL SOLUTIONS
Nova Biomedical Corp., Apr 28, 2006
K070068NOVA STATSENSOR CREATININE HOSPITAL METER, CONTROL SOLUTIONS,LINEARITY SOLUTIONS, AND TEST STRIPS
Nova Biomedical Corp., May 7, 2007