ClearanceBrief
Recall Z-1614-2008 ยท IYE Accelerator, Linear, Medical

Med Tec Inc dba CIVCO Medical Solutions

Fiducial Markers (0.9 x 3 mm) Soft tissue gold markers, Medical charged-particle radiation therapy system; Part Number: MT-NW-887-809, CIVCO Medical Solutions (MEDTEC), Orange City, IA 51041.

DateMay 21, 2008 (initiated Feb 28, 2008)
StatusTerminated
ReasonMislabeled: Fiducial markers were incorrectly labeled. They are marked (0.9x3mm cylindrical) Soft tissue gold markers, but contain (2mm spheres) Bone gold markers.
Root causeLabeling mix-ups
ActionConsignees were notified via telephone beginning 2/28/08 and a Recall Notification letter sent on 2/29/08. Users were requested to return any affected products for a replacement. They were also requested to return a recall notification form indicating the amount of affected product in their inventory. For additional information, call 800-842-8688.
Quantity33/3-marker pouches
DistributionNationwide Distribution including states of WA, PA, DE, AL, CT, WI, FL, TX, MN, and IL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071614FIDUCIAL MARKERS
Civco, Sep 11, 2007