Recall Z-1349-2008 ยท LLZ System, Image Processing, Radiological

AGFA Corp.

Agfa IMPAX CardioVascular Suite Results Management Nuclear Reporting Module, Outbound Report, Model/Catalog No.: L7W1P, Software Versions: 2.04.17, 2.04.19 and 2.04.20, AGFA HealthCare, Westerly, RI

DateMar 31, 2008 (initiated Feb 22, 2008)
StatusTerminated
ReasonReporting Differences: The Outbound Report software produces an Outbound Report (text version) which does not contain all the clinical content elements present in the PDF version of the report, also generated on the Nuclear Cardiology Reporting Module.
Root causeSoftware design
ActionAll five (5) sites utilizing this product were notified about the issue directly via telephone. An "Urgent Safety Notice" was sent via FED-EX to all the sites using the product on February 22, 2008. The letter describes the potential issue and mitigation. The recalling firm will also visit each site and make configuration changes to resolve this issue.
Quantity5 units
DistributionNationwide to five hospitals in DC, NY, OH, TN, and VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050228AGFA HEARTLAB CARDIOVASCULAR
Heartlab, Inc., Feb 18, 2005