Recall Z-1459-2008 ยท IYE Accelerator, Linear, Medical

Varian Medical Systems Oncology Systems

High-Definition 120 MultiLeaf Collimator, Medical charged-particle radiation therapy system; Software Version 7.0; Product Code: H56; Accessory to Medical Linear Accelerator.

DateAug 2, 2008 (initiated Jan 30, 2008)
StatusTerminated
ReasonDose Calculation Error: Software anomaly may result in failure of a MLC leaf to reach planned position, potentially resulting in misadministration of dose to a patient.
Root causeSoftware design
ActionConsignees were notified by a Medical Device Notification letter sent on 01/29/2008. The letter advised users to cease use of the device until a corrected version of the software is available. A representative will contact users to upgrade the application once the revised software has been validated and released for distribution.
Quantity3 units subject to correction
DistributionNationwide Distribution including states of NC, FL and CA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071992HIGH-DEFINITION 120 MULTILEAF COLLIMATOR (HD120 MLC)
Varian Medical Systems, Aug 15, 2007