ClearanceBrief
Recall Z-1467-2008 ยท CBK Ventilator, Continuous, Facility Use

Cardinal Health 203, Inc

Replacement Kit, Power PCBA with DC power cord. REF 11590. Pulmonetic Systems, 17400 Medina Rd., Suite 100, Minneapolis, MN 55447, USA.

DateAug 13, 2008 (initiated Feb 22, 2008)
StatusTerminated
ReasonThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Root causeDevice Design
ActionAn Urgent Medical Device Recall letter dated February 22, 2008 was sent to ventilator consignees/owners and distributors. The letter describes the issue, identifies the affected devices, states that the devices will be corrected by Cardinal Health and provides direction to verify ventilator operation and audible alarm function daily. Service Centers were provided information to identify affected printed circuit board assembly replacement kits' lot numbers and asked to return them. An "Update" was mailed 03/04/2008 to identify additional ventilator serial numbers that also need correction.
Quantity503 (269 US, 234 OUS)
DistributionWorldwide Distribution --- USA including Puerto Rico and countries of Algeria, Argentina, Australia, Austria, Bahamas, Belgium, Brazil, Canada, Chile, China Peoples Republic, Colombia, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Israel, Italy, Japan, Malaysia, Mexico, New
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002881VENTILATOR, CONTINUOUS (RESPIRATOR), MODEL LTV 1000
Pulmonetic Systems, Inc., Oct 11, 2000
K010508PASCO MIC AND MIC/ID PANELS
Pasco Laboratories, Inc., Apr 23, 2001
K040540MODIFICATION TO LTV 1000 VENTILATOR
Pulmonetic Systems, Inc., Apr 29, 2004
K981371LTV 1000 VENTILATOR
Pulmonetic Systems, Inc., Oct 30, 1998
K984056LTV 1000
Pulmonetic Systems, Inc., Apr 12, 1999