ClearanceBrief
Recall Z-1181-2008 ยท MEA Pump, Infusion, Pca

Hospira Inc.

Hospira LifeCare PCA Label Utility, Version 1.0 CD. The installation/user guide is entitled "LifeCare PCA Barcode Label Utility Program." This software program is designed to allow entry of each drug concentration into a formulation library on a pharmacy computer for standardized printing of barcode labels for compounded drug product vials.

DateAug 6, 2008 (initiated Jan 3, 2008)
StatusTerminated
ReasonIncorrect dosage labels created: Customers have experienced independent changes in drug concentration (mg/mL to mcg/mL and vice versa) and expiration dates, and the addition of trailing zeros to the barcodes.
Root causeSoftware design
ActionHospira sent letters dated 1/3/08 to the immediate recipients of the product via FedEx overnight letter on the same date. The letters informed the accounts of the reports of independent changes in drug concentration while editing a saved formulation using the Hospira LifeCare PCA Label Utility Software CD that was provided to some customers in a Hospira MedNet Drug Library Creation Team Introductory Kit-LifeCare PCA kit. The accounts were requested to remove the utility from any PC in their facility on which it is currently installed, destroy the CD and sign and return the enclosed removal v
Quantity127 CDs
DistributionNationwide Distribution and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.