Recall Z-1563-2008 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

GE Healthcare

GE Healthcare Precision RXi 23A/ 32A Remote X-Ray Imaging R&F System. It is a remote analog R&F system consisting of a tilting table with integrated spot film device, 65 kW pulsing generator and high heat capacity X-Ray tube. The system can be configured to meet a wide variety of exams for various clinical applications with ether 23 cm image intensifier with 512x512 Image Chain or 32 cm Image int

DateAug 24, 2008 (initiated Dec 18, 2007)
StatusTerminated
ReasonGE Healthcare Precision RXi tilt joystick became stuck in the Trendelenburg direction causing the patient to slide off the table. Also recall conducted to investigate the possibility of accidentally activation of the remote console joystick where a fatal accident occurred on a non-GE Remote R&F System. .
Root causeDevice Design
ActionTwo letters "GE Product Safety Notification" dated 12/18/07 and 3/7/08, addressed to Hospital Administrators, Manager of Radiology and Radiologists, were sent to each consignee. Each letter addressed a separate safety issue. Letters described the affected product, safety instructions, corrective action and contact information. The joystick can become stuck causing unintentional motion of the tube. Field Modification Instructions (FMI 10802/10803) have been issued to the field to install additional warning labels on the patient table, install a protector for the console joystick and inspect
Quantity66
Distribution(USA) AK CO IL LA MA MI MS NC NJ NY PA SC TN UT VA WA WI (OUSA) BR CA MX AU C3 KR PH TW AE AT BE BF BG CM CY CZ DE ES FI FR GB GE GR GY HR HU IT LB LV MT NA NG PL PT RO RU SA SD SE S
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041605PRECISION RXI ANALOG X-RAY SYSTEM
General Medical Merate S.P.A, Jun 30, 2004