ClearanceBrief
Recall Z-1237-2008 ยท DQA Oximeter

International Technidyne Corp.

International Technidyne Corp. (ITC) AVOXimeter 1000 Oximeter, Catalog Numbers: AVON1000-110 and AVOX-1000-220

DateAug 6, 2008 (initiated Feb 13, 2008)
StatusTerminated
ReasonIncorrect reading: ITC discovered that instruments built or repaired since approximately April 2007 were inadvertently left with the diagnostic mode enabled when shipped to customers.
Root causeProcess design
ActionITC sent Urgent - Medical Device Field Recall - ITC No. 08-001 letters dated 2/08/2008 to hospitals and/or medical facilities identified as having affected instruments. The letter instructed users not to use the instrument if the diagnostic mode is enabled. ITC will reset affected devices and return them to the customers.
Quantity7 units
DistributionWorldwide Distribution: USA, United Arab Emirates, Malaysia, Saudi Arabia, India, Germany, United Kingdom, Greece, Romania, Singapore, ROC, Pakistan, Netherlands, Korea, France, Belgium, Switzerland, Canada, Thailand, Slovenia, Turkey and Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K922075AVOXIMETER
A-Vox Systems, Inc., Nov 4, 1992