ClearanceBrief
Recall Z-1319-2008 ยท GTQ Antistreptolysin - Titer/Streptolysin O Reagent

BD Diagnostic Systems / Lee Laboratories

DIFCO Laboratories, FA Streptococcus Group A, 5mL, Catalog Number: 223181, Lot number: 6122656, Exp: 20090705, in vitro diagnostic.

DateSep 11, 2008 (initiated Aug 8, 2007)
StatusTerminated
ReasonIncorrect result: The product exhibited decreased fluorescence when testing with the homologous control organism.
Root causeNonconforming Material/Component
ActionBecton Dickinson (BD) Diagnostic Systems sent an Urgent product recall letter, dated August 2007, to all consignees, directing them to discard this lot of product. BD instructed distributors to discontinue distribution and discard any remaining product in inventory. BD requested the distributors provide an Excel file or a customer list containing the name, address and telephone number of end-users who were shipped the product. BD provided alternative testing methods. A response form was included to verify receipt of notification.
Quantity26 units
DistributionNationwide Distribution and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.