ClearanceBrief
Recall Z-2036-2008 ยท LDD Dc-Defibrillator, Low-Energy, (Including Paddles)

Physio Control, Inc.

Internal Defibrillation Electrodes: PN: 805355-10 thru 14 and 805355-020 thru -024.

DateSep 16, 2008 (initiated Jan 29, 2008)
StatusTerminated
ReasonBased on testing, sterilization methods described in labeling may be ineffective and may cause damage or corrosion to the paddles or handles. If handles and/or paddles are not sterilized adequately and are used on a patient, there is a risk of infection. If handle discharge button is damaged during sterilization, defibrillation therapy may be prevented.
Root causeLabeling design
ActionStarting 2/5/08, Physio-Control sent letters to customers which notifies them not to use Flash Gravity Steam, 12/88 Ethylene Oxide, STERRAD 100 machine, or Gravity Steam to sterilize. The firm recommends the use of 100% EO, STERRAD 100S machine, or Prevacuum steam for sterilization. The letter also recommends that after sterilization, the handles and paddles should be examined carefully for damage. If damaged, the product should be removed from use. The firm will follow up with a revised Sterilization Guideline document and requests posting the letter in the sterilization department in the
DistributionDistributed to hospitals and medical centers. Nationwide, DOD, and international: ARGENTINA, BAHRAIN, BANGLADESH, BERMUDA, BOLIVIA, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, EGYPT, EL SALVADOR, FRANCE, GERMANY, GREECE, GUATEMALA, GUYANA, HONG KONG, INDIA, IRAN, ISRAE
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.