ClearanceBrief
Recall Z-0969-2008 ยท JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented

Stryker Howmedica Osteonics Corp.

Stryker Trial Head xx-Long (+16mm) Neck Length 28mm Diameter Sterile, Forged Vitallium Alloy; Product Number: 6264-8-528; Howmedica Osteonics Corp. Styrker Ireland; Carrigwohill Inustrial Estate; Carrigtwohill county Cork, Ireland. Made in Ireland.

DateFeb 7, 2008 (initiated Nov 2, 2007)
StatusTerminated
ReasonFracture of Stem Neck - The firm recently performed laboratory testing indicates that there may be a potential increased risk of stem neck fractures when a V40 +16mm offset femoral head is mated with certain Accolade TMZF femoral stems in patients classified as overweight or obese.
Root causeOther
ActionStryker Orthopaedics sent an Urgent Product Recall notification letters on November 9, 2007, by Federal Express with return receipt. The letter informed users of potential increased risk of stem neck fracture, when a V40 + 16mm offset femoral head is mated with certain Accolade TMZF femoral stems in patients classified as overweight or obese. The firm has asked that inventory be examined, removed and held in quarantine. A Stryker representative will contact the medical facilities to retrieve the product. If a representative has not contacted the facility to date, it is suggested that they
Quantity5,599 total units
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.