ClearanceBrief
Recall Z-1170-2008 ยท MEH Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate

Stryker Howmedica Osteonics Corp.

Trident Hemispherical Cluster; 42mm - 74 mm Arc Deposited; Hydroxylapatite Coated; Multiholed; Use with Size A insert; hip prosthesis component, Stryker Orthopaedics, Howmedica Osteonics Corp, Mahwah, NJ 07430

DateJun 12, 2008 (initiated Jan 21, 2008)
StatusTerminated
ReasonForeign material: Some of the parts tested exceeded Stryker Orthopaedics internal acceptance criteria for manufacturing residuals.
Root causeProcess design
ActionRecall notification letters were sent to Stryker Branches/agencies, OR Supervisors and Chief of Orthopaedics on 1/21/08. A Patient information sheet was sent on 2/4/08 to surgeons and hospitals. Per call with Center 2/4/08, recall letter to be revised and RES updated upon receipt. The letter was revised and sent on February 28, 2008 to include Trident PSL Acetabular shells. This letter was sent to Risk Management at hospitals and included the scope of the recall , the potential hazard and recommendation to physicians to monitor patients consistent with care for those receiving total hip re
Quantity129,312 total shells, all Trident Hemispherical types
DistributionNationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013676TRIDENT HEMISPHERICAL ACETABULAR SHELLS
Howmedica Osteonics Corp., Dec 5, 2001