Recall Z-1169-2008 ยท MEH Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate
Stryker Howmedica Osteonics Corp.
Trident Hemispherical Acetabular Shells HA Solid BK 42mm- 74mm; Arc Deposited; Hydroxylapatite Coated; Multiholed; Use with Size A insert; hip prosthesis component, Stryker Orthopaedics, Howmedica Osteonics Corp, Mahwah, NJ 07430
| Date | Jun 12, 2008 (initiated Jan 21, 2008) |
| Status | Terminated |
| Reason | Foreign material: Some of the parts tested exceeded Stryker Orthopaedics internal acceptance criteria for manufacturing residuals. |
| Root cause | Process design |
| Action | Recall notification letters were sent to Stryker Branches/agencies, OR Supervisors and Chief of Orthopaedics on 1/21/08. A Patient information sheet was sent on 2/4/08 to surgeons and hospitals. Per call with Center 2/4/08, recall letter to be revised and RES updated upon receipt. The letter was revised and sent on February 28, 2008 to include Trident PSL Acetabular shells. This letter was sent to Risk Management at hospitals and included the scope of the recall , the potential hazard and recommendation to physicians to monitor patients consistent with care for those receiving total hip re |
| Quantity | 129,312 total shells, all Trident hemispherical types |
| Distribution | Nationwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K013676 | TRIDENT HEMISPHERICAL ACETABULAR SHELLS Howmedica Osteonics Corp., Dec 5, 2001 |