ClearanceBrief
Recall Z-1053-2008 ยท FMZ Incubator, Neonatal

Draeger Medical, Inc.

Draeger Globe Trotter IC, Catalog Number: MU20509, Draeger Medical, Inc., Telford, PA 18969

DateMar 18, 2008 (initiated Jan 31, 2008)
StatusTerminated
ReasonHeating failure- The power board which controls the incubator heater may not regulate the temperature properly resulting in a high temperature alarm and a loss of temperature control within the patient compartment.
Root causeNonconforming Material/Component
ActionThe firm notified consignees on 2/1/08, by an Important Safety Notice. The letter instructed users to remove any units from use until a Draeger Service Representative can replace the power board. A Service Representative will be in contact to schedule a service visit. For questions concerning this recall please contact 215-660-2418 or Anbu.arasu@draeger.com.
Quantity4 units
DistributionWorldwide Distribution-- USA including states MO, OH, and PR, and countries of Australia, Canada, Columbia, Israel, UAE, France, China, Italy, Denmark, Switzerland, and UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K001019TI500 GLOBE-TROTTER TRANSPORT SYSTEM
Hill-Rom Air-Shields, Apr 27, 2000