ClearanceBrief
Recall Z-0961-2008 ยท LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief

Advanced Bionics Corp

Advanced Bionics Precision Implantable Pulse Generator Kit, Model Number SC-1100 with firmware version prior to Revision 3.02, Advanced Bionics Corportation, Sylmar, CA 91342

DateFeb 27, 2008 (initiated Oct 5, 2007)
StatusTerminated
ReasonIncorrect Data -- Corruption of internal memory component results in an inability for the physician to reprogram the IPG with firmware version prior to Revision 3.02. When this occurs, the IPG will report an error code of "10h0" or "00h0" through the Remote Control. Under this condition, the IPG will cease to log in some data that could be used for informational purposes. ** It should be noted t
Root causeDevice Design
ActionAdvanced Bionics sent an Important Notification letter on October 5, 2007, via FED-Ex or US Postal Service Certified mail. (International notification activity is in progress.) The recall letter advises that the firm has become aware of an issue with Precision Implantable Pulse Generators (IPG model numbers SC-1100, Sc1110, and ONS-1110) manufactured prior to July 2007. The letter describes the issue in detail. The firm is requesting that customer immediately locate the affected IPGs and set them aside so that they are not used. A Boston Scientific/Advanced Bionics sales representative will
QuantityOver 12,000 units
DistributionClass II Recall - Worldwide Distribution --- including USA and countries of EU, Canada, Australia and Israel.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.